Zimmer Biomet, Inc.: Medical Device Recall in 2018 - (Recall #: Z-2367-2018)
See the recall detail below. You can also see other recalls from the same firm in 2018.
Zimmer¿ M/L Taper Hip Prosthesis With Kinective¿ Technology Cementless, Size 12.5; 00-7713-012-00
Class II
M/L Taper Hip Prosthesis and M/L Taper Hip Prosthesis with Kinectiv Technology manufactured prior to October 2015 are being recalled due to the potential for black residue to be observed within the polyethylene bag containing the implant. The black residue is a combination of the titanium plasma spray used to create the porous surface of the implant and polyethylene bag containing the implant. The recalling firm states the residue is biocompatible. Risks include delay of surgery while locating a replacement device, and surgical intervention due to wear caused by debris.
61741690 62808128 62831371 63068944 63090038 63094522 63128589
Products were distributed solely to Japan.
Voluntary: Firm initiated