Zimmer Biomet, Inc.: Medical Device Recall in 2018 - (Recall #: Z-2373-2018)
See the recall detail below. You can also see other recalls from the same firm in 2018.
Zimmer¿ M/L Taper Hip Prosthesis With Kinective¿ Technology Cementless, Size 22.5; 00-7713-022-00
Class II
M/L Taper Hip Prosthesis and M/L Taper Hip Prosthesis with Kinectiv Technology manufactured prior to October 2015 are being recalled due to the potential for black residue to be observed within the polyethylene bag containing the implant. The black residue is a combination of the titanium plasma spray used to create the porous surface of the implant and polyethylene bag containing the implant. The recalling firm states the residue is biocompatible. Risks include delay of surgery while locating a replacement device, and surgical intervention due to wear caused by debris.
00111778 60940196 61025425 61063904 61097911 61120383 61134445 61619526 62420837 62798037
Products were distributed solely to Japan.
Voluntary: Firm initiated