Zimmer Biomet, Inc.: Medical Device Recall in 2018 - (Recall #: Z-2375-2018)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2018.

Data Source: FDA.
Product Description:

Zimmer¿ M/L Taper Hip Prosthesis With Kinective¿ Technology Cementless, Size 9; 00-7713-009-00

Product Classification:

Class II

Date Initiated: May 30, 2018
Date Posted: July 11, 2018
Recall Number: Z-2375-2018
Event ID: 80188
Reason for Recall:

M/L Taper Hip Prosthesis and M/L Taper Hip Prosthesis with Kinectiv Technology manufactured prior to October 2015 are being recalled due to the potential for black residue to be observed within the polyethylene bag containing the implant. The black residue is a combination of the titanium plasma spray used to create the porous surface of the implant and polyethylene bag containing the implant. The recalling firm states the residue is biocompatible. Risks include delay of surgery while locating a replacement device, and surgical intervention due to wear caused by debris.

Status: Terminated
Product Quantity: 116
Code Information:

62350575 62647167 62787292 63044896 63044897 63068905 63076554 63120523 63161387 63174034

Distribution Pattern:

Products were distributed solely to Japan.

Voluntary or Mandated:

Voluntary: Firm initiated