Zimmer Biomet, Inc.: Medical Device Recall in 2019 - (Recall #: Z-0525-2020)
Updated on March 25, 2026.
See the recall detail below. You can also see other recalls from the same firm in 2019.
Data Source: FDA.
Product Description:
Comprehensive RS Cleat CAP, Item No. 110300
Product Classification:
Class II
Date Initiated: August 21, 2019
Date Posted: December 4, 2019
Recall Number: Z-0525-2020
Event ID: 83783
Reason for Recall:
Elevated levels of bacterial endotoxin and residual debris remain on the devices due to cleaning issue.
Status: Terminated
Product Quantity: 219988 units (total)
Code Information:
126590 126600 502890 502900 502910 012080 941080
Distribution Pattern:
US Nationwide distribution and countries of Argentina, Australia, Brazil, Canada, Chile, Colombia, Costa Rica, Ecuador, El Salvador, Italy, Netherlands, India, Japan, Korea, Malaysia, Mexico, New Zealand, Panama, Singapore, Taiwan, Thailand, Trinidad and Tobago, and Venezuela.
Voluntary or Mandated:
Voluntary: Firm initiated