Zimmer CAS: Medical Device Recall in 2014 - (Recall #: Z-0495-2015)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2014.

Data Source: FDA.
Product Description:

PSI Knee System- Prosthesis, knee patellofemorotibial, semi-constrained, cemented, polymer/meal/polymer The Zimmer PSI Knee System is indicated as an orthopedic instrument system to assist in the positioning of knee replacement components.

Product Classification:

Class II

Date Initiated: November 3, 2014
Date Posted: December 10, 2014
Recall Number: Z-0495-2015
Event ID: 69630
Reason for Recall:

Internal discovery found that the MRI scans in some PSI knee guides involved an additional rotation adjustments that could potentially lead to misalignment due to a design flaw.

Status: Terminated
Product Quantity: 15
Code Information:

BKL0398L 20-8070-002-01; DHA0645L 20-8070-002-01; DHA0646R 20-8070-002-02; ECO0516L 20-8070-002-01; GHE0651R 20-8070-002-02; KAD0595L 20-8070-002-01; KWI0523R 20-8070-002-02; MHA0483L 20-8070-002-01; MHA0484R 20-8070-002-02; MKE0659R 20-8070-002-02; MMA0728R 20-8070-002-02; SSC0782L 20-8070-001-01; TCA0582L 20-8070-002-01; THO0803L 20-8070-002-01; TMA0760L 20-8070-002-01; Expiration dates 5/8/14-12/31/15

Distribution Pattern:

US Distribution in states of: AZ, CO, IL, OR, and WA.

Voluntary or Mandated:

Voluntary: Firm initiated