Zimmer Gmbh: Medical Device Recall in 2015 - (Recall #: Z-1523-2015)
Updated on March 25, 2026.
See the recall detail below. You can also see other recalls from the same firm in 2015.
Data Source: FDA.
Product Description:
Anatomical Shoulder Handle for Rasp Item: 01.04233.000 Used during total joint arthroplasty of the shoulder.
Product Classification:
Class II
Date Initiated: March 18, 2015
Date Posted: May 6, 2015
Recall Number: Z-1523-2015
Event ID: 70993
Reason for Recall:
Complaints of difficulties to attach the Anatomical Shoulder Rasps (high resistance) or, once attached, due to difficulties to remove the handle (seizing up of the two components). This has the potential for delay in surgery.
Status: Terminated
Product Quantity: 30 units
Code Information:
Lot Number(s): 13870129 & 13878179.
Distribution Pattern:
Distributed to the states of IN, GA, WI, MO, PA, KS, TN, and NY.
Voluntary or Mandated:
Voluntary: Firm initiated