Zimmer Gmbh: Medical Device Recall in 2016 - (Recall #: Z-2003-2016)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2016.

Data Source: FDA.
Product Description:

Dome Centric component (Reference 01.04227.005) from the Anatomical Shoulder (AS) System 2.0. The Dome Centric connects the Humeral Stem to the Humeral Head and allows setting an inclination and retroversion positioning. The Anatomical Shoulder System is intended for long-term implantation into the human shoulder joint in primary or revision, total or hemi shoulder arthroplasty. The system is intended to relieve pain and restore function in patients with adequate bone stock to support the prosthesis.

Product Classification:

Class II

Date Initiated: May 17, 2016
Date Posted: June 22, 2016
Recall Number: Z-2003-2016
Event ID: 74206
Reason for Recall:

In some cases it has been difficult or not possible to disassemble the adjusted Dome Centric from the AS Humeral Rasp after initial positioning of the AS Humeral Trial Head which resulted in the whole construct of the Dome Centric with the still assembled AS Humeral Rasp taken out of the humeral canal. This could cause a delay in surgery time or the surgeon could decide to close the wound without finishing the surgery, increasing the infection risk or second use of anesthesia necessary.

Status: Terminated
Product Quantity: 177
Code Information:

Item Number: 01.04227.005 Lot Numbers: 2792457 2794578 2812029 2817212 2844463 2794576 2794579 2815118 2817213 2847568 2794577 2794580 2815119 2832862

Distribution Pattern:

Nationwide Distribution to AZ, CA, CO, GA, IL, KS, KY, MA, MD, MO, OH, PA, TN, and TX.

Voluntary or Mandated:

Voluntary: Firm initiated