Zimmer, Inc.: Medical Device Recall in 2012 - (Recall #: Z-0296-2013)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2012.

Data Source: FDA.
Product Description:

ATS 3000 Automatic Tourniquet System w/HOSES AND LOP SENSOR Intended to be used by qualified medical professionals to temporarily occlude blood flow in a patient's extremities during surgical procedures on those extremities.

Product Classification:

Class II

Date Initiated: September 5, 2012
Date Posted: November 21, 2012
Recall Number: Z-0296-2013
Event ID: 63458
Reason for Recall:

Zimmer is conducting a voluntary removal of some Zimmer A.T.S. 3000 tourniquets after receiving complaints of both out-of-box failures and failures during surgical procedures on some of the affected units. During failure the unit will alarm, lock into a non-operational mode, and the display screen will go blank. Based on investigation, this problem appears to be due to a counterfeit chip on the

Status: Terminated
Product Quantity: 204
Code Information:

Catalog 60-3000-101-00, Serial # 3005IACJ, 3006JAEP, 3007CADR, 3007EACT, 3007EAKT, 3007GAGR, 3007HABK, 3007KAAR, 3007KAET, 3008CADE, 3008HAHC, 3008LACK, 3009AACN, 3009HABA, 3009LAEF, 3010CABH, 3010DAHG, 3010GAAC, 3010GAAJ, 3011LABL, 3011LABN, 3011LABQ, 3011LACC, 3011LACE, 3011LACF, 3011LACG, 3011LACH, 3011LACJ, 301 1LACK, 3011LACL, 3011LACM, 3011LACN, 3011LACP, 3011LACQ, 3011LACR, 3011LACS, 3011LACT, 3011LACU, 3011LACW, 3011LADA, 3011LADE, 3011LADF, 3011LADG, 3011LADH, 3011LADJ, 3011LADK, 3011LADM, 3011LADP, 3011LADQ, 3011LADR, 3011LADS, 3011LADT, 3011LADU, 3011LADW, 3011LAEA, 3011LAEB, 3011LAEC, 3011LAED, 3011LAEE, 3011LAEF, 3011LAEH, 3012AAAG, 3012AAAK, 3012AAAM, 3012AAAN, 3012AAAP, 3012AAAQ, 3012AAAR, 3012AAAS, 3012AAAT, 3012AAAU, 3012AAAW, 3012AABA, 3012AABB, 3012AABC, 3012AABE, 3012AABG, 3012AABH, 3012AABK, 3012AABL, 3012AABM, 3012AABN, 3012AABQ, 3012AABR, 3012AABS, 3012AABU, 3012AACA, 3012AACB, 3012AACC, 3012AACD, 3012AACE, 3012AACH, 3012AACJ, 3012AACM, 3012AACN, 3012AACP, 3012AACQ, 3012AACR, 3012AACS, 3012AACT, 3012AACW, 3012AADA, 3012AADB, 3012AADC, 3012AADD, 3012AADE, 3012AADF, 3012AADG, 3012BAAA, 3012BAAB, 3012BAAC, 3012BAAD, 3012BAAE, 3012BAAF, 3012BAAG, 3012BAAH, 3012BAAJ, 3012BAAK, 3012BAAL, 3012BAAM, 3012BAAN, 3012BAAP, 3012BAAQ, 3012BAAR, 3012BAAS, 3012BAAT, 3012BAAU, 3012BAAW, 3012BABA, 3012BABB, 3012BABC, 3012BABD, 3012BABE, 3012BABF, 3012BABG, 3012BABH, 3012BABJ, 3012BABK, 3012BABL, 3012BABM, 3012BABN, 3012BABP, 3012BABQ, 3012BABR, 3012BABS, 3012BABT, 3012BABU, 3012BABW, 3012BACE, 3012BACG, 3012BACH, 3012BACJ, 3012BACK, 3012BACL, 3012BACM, 3012BACN, 3012BACP, 3012BACQ, 3012BACR, 3012BACS, 3012BACT, 3012BACU, 3012BACW, 3012BADA, 3012BADB, 3012BADC, 3012BADD, 3012BADG, 3012BADH, 3012BADJ, 3012BADK, 3012BADL, 3012BADN, 3012BADP, 3012BADR, 3012BADS, 3012BADU, 3012BADW, 3012BAEA, 3012BAEB, 3012BAEC, 3012BAEH, 3012BAEJ, 3012BAEL, 3012BAEM, 3012BAEN, 3012BAEQ, 3012BAER French Catalog 60-3000 301-02 Serial # 3012 AAAE German Catalog 60-300060-301-03, Serial # 3012AAAF Spanish Catalog 60-3000-301-11, Serial # 3012AAAB, 3012AAAC, 3012AACG, 3012BACA, 3012BACB, 3012BACC, 3012BACD Dutch 60-3000-301-12, Serial # 3012AAAA, 3012AAAD, 3012EACA

Distribution Pattern:

Worldwide Distribution - USA (nationwide) and Internationally to Belgium, Czech Republic, Egypt, France, Germany, Morocco, Netherlands, Romania, Saudi Arabia, Spain, Switzerland, Turkey, and UAE.

Voluntary or Mandated:

Voluntary: Firm initiated