Zimmer, Inc.: Medical Device Recall in 2013 - (Recall #: Z-0776-2013)
See the recall detail below. You can also see other recalls from the same firm in 2013.
Trilogy Acetabular AB Shells with Holes. The Trilogy Acetabular System is indicated for either cemented or uncemented use in individuals undergoing surgery for rehabilitating hip damage.
Class II
Some units of Trilogy shells manufactured since March 2009 may be below the lower range of the pore size specification (as low as 28% vs. 30% porosity specification). On 10/31/2012, the firm expanded the recall action to include lots of Uni-Hole Trilogy Acetabular System shells.
62054788 61901417 61984072 61901423 61901426 61901427 61901428 61901435 61901437 61278096
Worldwide Distribution-USA (nationwide) and the countries of Denmark, Germany, Greece, Hungary, Ireland, Italy, Netherlands, Spain, Sweden, UK, Poland, France, Australia, Chile, Canada, India, Japan, New Zealand, Peru, South Africa, Colombia, Ukraine, and UAE.
Voluntary: Firm initiated