Zimmer, Inc.: Medical Device Recall in 2013 - (Recall #: Z-1998-2013)
See the recall detail below. You can also see other recalls from the same firm in 2013.
00597001101 Femur CR PRECOAT FEM COMP SIZE AML, Rx, Sterile; 00597001102 Femur CR PRECOAT FEM COMP SIZE AMR, Rx, Sterile; 00597001201 Femur CR PRECOAT FEM COMP SIZE BML, Rx, Sterile; 00597001202 Femur CR PRECOAT FEM COMP SIZE BMR, Rx, Sterile. Used in total knee arthroplasty and indicated for patients with severe pain.
Class II
Surgeons are implanting components of the NexGen Cruciate Retaining (CR) Complete Knee System in combinations that are not approved (micro with standard). Includes CR Micro Femurs, Articular Surface, Patellae, instruments and provisionals combined with standard components. A March 2012 recall reiterated the proper combination labeling, but the mis match problem continues. Zimmer is removing the d
REF 00597001101, all codes
Worldwide Distribution-USA (nationwide) and the country of Japan.
Voluntary: Firm initiated