Zimmer, Inc.: Medical Device Recall in 2013 - (Recall #: Z-2009-2013)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2013.

Data Source: FDA.
Product Description:

00597001303 Femur CR PRC FEM COMP SIZE CML Rx, Sterile; 00597001304 Femur CR PRC FEM COMP SIZE CMR Rx, Sterile; 00597001403 Femur CR PRC FEM COMP SIZE DML Rx, Sterile; 00597001404 Femur CR PRC FEM COMP SIZE DMR Rx, Sterile; 00597001503 Femur CR PRC FEM COMP SIZE EML Rx, Sterile; 00597001504 Femur CR PRC FEM COMP SIZE EMR Rx, Sterile; Used in total knee arthroplasty and indicated for patients with severe pain.

Product Classification:

Class II

Date Initiated: June 7, 2013
Date Posted: August 28, 2013
Recall Number: Z-2009-2013
Event ID: 65413
Reason for Recall:

Surgeons are implanting components of the NexGen Cruciate Retaining (CR) Complete Knee System in combinations that are not approved (micro with standard). Includes CR Micro Femurs, Articular Surface, Patellae, instruments and provisionals combined with standard components. A March 2012 recall reiterated the proper combination labeling, but the mis match problem continues. Zimmer is removing the d

Status: Terminated
Product Quantity: 192,355
Code Information:

all codes

Distribution Pattern:

Worldwide Distribution-USA (nationwide) and the country of Japan.

Voluntary or Mandated:

Voluntary: Firm initiated