Zimmer, Inc.: Medical Device Recall in 2013 - (Recall #: Z-2012-2013)
See the recall detail below. You can also see other recalls from the same firm in 2013.
00597201333 Femur CR POR SURF HDN FEM CO-NID CML, Rx, Sterile; 00597201334 Femur CR POR SURF HDN FEM CO-NID CMR, Rx, Sterile; 00597201433 Femur CR POR SURF HDN FEM CO-NID DML, Rx, Sterile; 00597201434 Femur CR POR SURF HDN FEM CO-NID DMR, Rx, Sterile; 00597201533 Femur CR POR SURF HDN FEM CO-NID EML, Rx, Sterile; 00597201534 Femur CR POR SURF HDN FEM CO-NID EMR, Rx, Sterile; Used in total knee arthroplasty and indicated for patients with severe pain.
Class II
Surgeons are implanting components of the NexGen Cruciate Retaining (CR) Complete Knee System in combinations that are not approved (micro with standard). Includes CR Micro Femurs, Articular Surface, Patellae, instruments and provisionals combined with standard components. A March 2012 recall reiterated the proper combination labeling, but the mis match problem continues. Zimmer is removing the d
all codes
Worldwide Distribution-USA (nationwide) and the country of Japan.
Voluntary: Firm initiated