Zimmer, Inc.: Medical Device Recall in 2013 - (Recall #: Z-2014-2013)
See the recall detail below. You can also see other recalls from the same firm in 2013.
00587806126 Patella MICRO POROUS PAT 26MM X 10MM, Rx, Sterile; 00587806129 Patella MICRO POROUS PAT 29MM X 10MM, Rx, Sterile; 00587806132 Patella MICRO POROUS PAT 32MM X 10MM, Rx, Sterile; 00587806135 Patella MICRO POROUS PAT 35MM X 10MM, Rx, Sterile; Used in total knee arthroplasty and indicated for patients with severe pain.
Class II
Surgeons are implanting components of the NexGen Cruciate Retaining (CR) Complete Knee System in combinations that are not approved (micro with standard). Includes CR Micro Femurs, Articular Surface, Patellae, instruments and provisionals combined with standard components. A March 2012 recall reiterated the proper combination labeling, but the mis match problem continues. Zimmer is removing the d
all codes
Worldwide Distribution-USA (nationwide) and the country of Japan.
Voluntary: Firm initiated