Zimmer, Inc.: Medical Device Recall in 2013 - (Recall #: Z-2020-2013)
See the recall detail below. You can also see other recalls from the same firm in 2013.
00597202009 Articular Surface CR ART SURF 12/PURPLE 09, Rx, Sterile; 00597202010 Articular Surface CR ART SURF 12/PURPLE 10, Rx, Sterile; 00597202012 Articular Surface CR ART SURF 12/PURPLE 12, Rx, Sterile; 00597202014 Articular Surface CR ART SURF 12/PURPLE 14, Rx, Sterile; 00597202017 Articular Surface CR ART SURF 12/PURPLE 17, Rx, Sterile; 00597202020 Articular Surface CR ART SURF 12/PURPLE 20, Rx, Sterile; 00597202023 Articular Surface CR ART SURF 12/PURPLE 23, Rx, Sterile; Used in total knee arthroplasty and indicated for patients with severe pain.
Class II
Surgeons are implanting components of the NexGen Cruciate Retaining (CR) Complete Knee System in combinations that are not approved (micro with standard). Includes CR Micro Femurs, Articular Surface, Patellae, instruments and provisionals combined with standard components. A March 2012 recall reiterated the proper combination labeling, but the mis match problem continues. Zimmer is removing the d
all codes
Worldwide Distribution-USA (nationwide) and the country of Japan.
Voluntary: Firm initiated