Zimmer, Inc.: Medical Device Recall in 2013 - (Recall #: Z-2021-2013)
See the recall detail below. You can also see other recalls from the same firm in 2013.
00597602010 Articular Surface AC ART SURF 12/PUR 10, Rx, Sterile; 00597602012 Articular Surface AC ART SURF 12/PUR 12, Rx, Sterile; 00597602014 Articular Surface AC ART SURF 12/PUR 14, Rx, Sterile; 00597602017 Articular Surface AC ART SURF 12/PUR 17, Rx, Sterile; 00597602020 Articular Surface AC ART SURF 12/PUR 20, Rx, Sterile; 00597602023 Articular Surface AC ART SURF 12/PUR 23, Rx, Sterile; Used in total knee arthroplasty and indicated for patients with severe pain.
Class II
Surgeons are implanting components of the NexGen Cruciate Retaining (CR) Complete Knee System in combinations that are not approved (micro with standard). Includes CR Micro Femurs, Articular Surface, Patellae, instruments and provisionals combined with standard components. A March 2012 recall reiterated the proper combination labeling, but the mis match problem continues. Zimmer is removing the d
all codes
Worldwide Distribution-USA (nationwide) and the country of Japan.
Voluntary: Firm initiated