Zimmer, Inc.: Medical Device Recall in 2013 - (Recall #: Z-2028-2013)
See the recall detail below. You can also see other recalls from the same firm in 2013.
00595204110 Articular Surface XLPE CR ART SURF 5-6/STRGRN 10, Rx, Sterile; 00595204112 Articular Surface XLPE CR ART SURF 5-6/STRGRN 12, Rx, Sterile; 00595204114 Articular Surface XLPE CR ART SURF 5-6/STRGRN 14, Rx, Sterile; 00595204117 Articular Surface XLPE CR ART SURF 5-6/STRGRN 17, Rx, Sterile; 00595204120 Articular Surface XLPE CR ART SURF 5-6/STRGRN 20, Rx, Sterile; Used in total knee arthroplasty and indicated for patients with severe pain.
Class II
Surgeons are implanting components of the NexGen Cruciate Retaining (CR) Complete Knee System in combinations that are not approved (micro with standard). Includes CR Micro Femurs, Articular Surface, Patellae, instruments and provisionals combined with standard components. A March 2012 recall reiterated the proper combination labeling, but the mis match problem continues. Zimmer is removing the d
all codes
Worldwide Distribution-USA (nationwide) and the country of Japan.
Voluntary: Firm initiated