Zimmer, Inc.: Medical Device Recall in 2013 - (Recall #: Z-2031-2013)
See the recall detail below. You can also see other recalls from the same firm in 2013.
90597004109 Articular Surface CR ART SURF AE56/STR GRN 09MM, Rx, Sterile; 90597004110 Articular Surface CR ART SURF AE56/STR GRN 10MM, Rx, Sterile; 90597004112 Articular Surface CR ART SURF AE56/STR GRN 12MM, Rx, Sterile; 90597004114 Articular Surface CR ART SURF AE56/STR GRN 14MM, Rx, Sterile; 90597004117 Articular Surface CR ART SURF AE56/STR GRN 17MM, Rx, Sterile; 90597004120 Articular Surface CR ART SURF AE56/STR GRN 20MM, Rx, Sterile; Used in total knee arthroplasty and indicated for patients with severe pain.
Class II
Surgeons are implanting components of the NexGen Cruciate Retaining (CR) Complete Knee System in combinations that are not approved (micro with standard). Includes CR Micro Femurs, Articular Surface, Patellae, instruments and provisionals combined with standard components. A March 2012 recall reiterated the proper combination labeling, but the mis match problem continues. Zimmer is removing the d
all codes
Worldwide Distribution-USA (nationwide) and the country of Japan.
Voluntary: Firm initiated