Zimmer, Inc.: Medical Device Recall in 2013 - (Recall #: Z-2032-2013)
See the recall detail below. You can also see other recalls from the same firm in 2013.
00597204109 Articular Surface CR ART SURF 56/STRIPE GRN 9, Rx, Sterile; 00597204110 Articular Surface CR ART SURF 56/STRIPE GRN 10, Rx, Sterile; 00597204112 Articular Surface CR ART SURF 56/STRIPE GRN 12, Rx, Sterile; 00597204114 Articular Surface CR ART SURF 56/STRIPE GRN 14, Rx, Sterile; 00597204117 Articular Surface CR ART SURF 56/STRIPE GRN 17, Rx, Sterile; 00597204120 Articular Surface CR ART SURF 56/STRIPE GRN 20, Rx, Sterile; 00597204123 Articular Surface CR ART SURF 56/STRIPE GRN 23, Rx, Sterile; Used in total knee arthroplasty and indicated for patients with severe pain.
Class II
Surgeons are implanting components of the NexGen Cruciate Retaining (CR) Complete Knee System in combinations that are not approved (micro with standard). Includes CR Micro Femurs, Articular Surface, Patellae, instruments and provisionals combined with standard components. A March 2012 recall reiterated the proper combination labeling, but the mis match problem continues. Zimmer is removing the d
all codes
Worldwide Distribution-USA (nationwide) and the country of Japan.
Voluntary: Firm initiated