Zimmer, Inc.: Medical Device Recall in 2013 - (Recall #: Z-2033-2013)
See the recall detail below. You can also see other recalls from the same firm in 2013.
00597604110 Articular Surface AC ART SURF 56/STRIPED GRN 10, Rx, Sterile; 00597604112 Articular Surface AC ART SURF 56/STRIPED GRN 12, Rx, Sterile; 00597604114 Articular Surface AC ART SURF 56/STRIPED GRN 14, Rx, Sterile; 00597604117 Articular Surface AC ART SURF 56/STRIPED GRN 17, Rx, Sterile; 00597604120 Articular Surface AC ART SURF 56/STRIPED GRN 20, Rx, Sterile; 00597604123 Articular Surface AC ART SURF 56/STRIPED GRN 23, Rx, Sterile; Used in total knee arthroplasty and indicated for patients with severe pain.
Class II
Surgeons are implanting components of the NexGen Cruciate Retaining (CR) Complete Knee System in combinations that are not approved (micro with standard). Includes CR Micro Femurs, Articular Surface, Patellae, instruments and provisionals combined with standard components. A March 2012 recall reiterated the proper combination labeling, but the mis match problem continues. Zimmer is removing the d
all codes
Worldwide Distribution-USA (nationwide) and the country of Japan.
Voluntary: Firm initiated