Zimmer, Inc.: Medical Device Recall in 2013 - (Recall #: Z-2038-2013)
See the recall detail below. You can also see other recalls from the same firm in 2013.
00597106126 Provisional MIC POROUS PAT PROV 26MMX10MM, Rx, Sterile; 00597106129 Provisional MIC POROUS PAT PROV 29MMX10MM, Rx, Sterile; 00597106132 Provisional MIC POROUS PAT PROV 32MMX10MM, Rx, Sterile; 00597106135 Provisional MIC POROUS PAT PROV 35MMX10MM, Rx, Sterile; Used in total knee arthroplasty and indicated for patients with severe pain.
Class II
Surgeons are implanting components of the NexGen Cruciate Retaining (CR) Complete Knee System in combinations that are not approved (micro with standard). Includes CR Micro Femurs, Articular Surface, Patellae, instruments and provisionals combined with standard components. A March 2012 recall reiterated the proper combination labeling, but the mis match problem continues. Zimmer is removing the d
all codes
Worldwide Distribution-USA (nationwide) and the country of Japan.
Voluntary: Firm initiated