Zimmer, Inc.: Medical Device Recall in 2013 - (Recall #: Z-2039-2013)
See the recall detail below. You can also see other recalls from the same firm in 2013.
00597502010 Provisional AC ART SURF PROV 12/PUR 10, Rx, Sterile; 00597502012 Provisional AC ART SURF PROV 12/PUR 12, Rx, Sterile; 00597502014 Provisional AC ART SURF PROV 12/PUR 14, Rx, Sterile; 00597502017 Provisional AC ART SURF PROV 12/PUR 17, Rx, Sterile; 00597502020 Provisional AC ART SURF PROV 12/PUR 20, Rx, Sterile; 00597502023 Provisional AC ART SURF PROV 12/PUR 23, Rx, Sterile; Used in total knee arthroplasty and indicated for patients with severe pain.
Class II
Surgeons are implanting components of the NexGen Cruciate Retaining (CR) Complete Knee System in combinations that are not approved (micro with standard). Includes CR Micro Femurs, Articular Surface, Patellae, instruments and provisionals combined with standard components. A March 2012 recall reiterated the proper combination labeling, but the mis match problem continues. Zimmer is removing the d
all codes
Worldwide Distribution-USA (nationwide) and the country of Japan.
Voluntary: Firm initiated