Zimmer, Inc.: Medical Device Recall in 2013 - (Recall #: Z-2040-2013)
See the recall detail below. You can also see other recalls from the same firm in 2013.
00597503110 Provisional AC ARTSURF PROV 34/STRIPE YEL10, Rx, Sterile; 00597503112 Provisional AC ARTSURF PROV 34/STRIPE YEL12, Rx, Sterile; 00597503114 Provisional AC ARTSURF PROV 34/STRIPE YEL14, Rx, Sterile; 00597503117 Provisional AC ARTSURF PROV 34/STRIPE YEL17, Rx, Sterile; 00597503120 Provisional AC ARTSURF PROV 34/STRIPE YEL 20, Rx, Sterile; 00597503123 Provisional AC ARTSURF PROV 34/STRIPE YEL 23, Rx, Sterile; Used in total knee arthroplasty and indicated for patients with severe pain.
Class II
Surgeons are implanting components of the NexGen Cruciate Retaining (CR) Complete Knee System in combinations that are not approved (micro with standard). Includes CR Micro Femurs, Articular Surface, Patellae, instruments and provisionals combined with standard components. A March 2012 recall reiterated the proper combination labeling, but the mis match problem continues. Zimmer is removing the d
all codes
Worldwide Distribution-USA (nationwide) and the country of Japan.
Voluntary: Firm initiated