Zimmer, Inc.: Medical Device Recall in 2013 - (Recall #: Z-2043-2013)
See the recall detail below. You can also see other recalls from the same firm in 2013.
00597103110 Provisional CR ART SURF PROV 34/STR YEL 10, Rx, Sterile; 00597103112 Provisional CR ART SURF PROV 34/STR YEL 12, Rx, Sterile; 00597103114 Provisional CR ART SURF PROV 34/STR YEL 14, Rx, Sterile; 00597103117 Provisional CR ART SURF PROV 34/STR YEL 17, Rx, Sterile; 00597103120 Provisional CR ART SURF PROV 34/STR YEL 20, Rx, Sterile; 00597103123 Provisional CR ART SURF PROV 34/STR YEL 23, Rx, Sterile; Used in total knee arthroplasty and indicated for patients with severe pain.
Class II
Surgeons are implanting components of the NexGen Cruciate Retaining (CR) Complete Knee System in combinations that are not approved (micro with standard). Includes CR Micro Femurs, Articular Surface, Patellae, instruments and provisionals combined with standard components. A March 2012 recall reiterated the proper combination labeling, but the mis match problem continues. Zimmer is removing the d
all codes
Worldwide Distribution-USA (nationwide) and the country of Japan.
Voluntary: Firm initiated