Zimmer, Inc.: Medical Device Recall in 2014 - (Recall #: Z-0589-2014)
Updated on March 25, 2026.
See the recall detail below. You can also see other recalls from the same firm in 2014.
Data Source: FDA.
Product Description:
ITST THREADED GUIDE PIN
Product Classification:
Class II
Date Initiated: November 5, 2013
Date Posted: January 8, 2014
Recall Number: Z-0589-2014
Event ID: 66499
Reason for Recall:
Based on a review of historic packaging validations, it was determined that a subset of product packaged at a Zimmer facility required testing to ensure that previously packaged product from this facility met the established requirements for sterile barrier integrity. This testing determined that a packaging tray configuration which consists of an inner and outer sterile tray, resulted in failure
Status: Terminated
Product Quantity: 42, 065 units total
Code Information:
Item Numbers: 47-2258-067-00 All Lots with an expiration date prior to July 24, 2023
Distribution Pattern:
Worldwide Distribution.
Voluntary or Mandated:
Voluntary: Firm initiated