Zimmer, Inc.: Medical Device Recall in 2014 - (Recall #: Z-1250-2014)
See the recall detail below. You can also see other recalls from the same firm in 2014.
MOST Options" System (MOST) HINGED TIB INS SZ 1 16MM HINGED TIB INS SZ 4 6MM M HINGED TIB INS SZ 4 9MM M HINGED TIB INS SZ 4 16MM MOST PROX TIB INS 6MM
Class II
A complaint was received for a discolored All Poly Tibia. Root cause analysis found that the discoloration was the result of free radical oxidation. The oxidation process was advanced prematurely as a result of the machined component being stored in a transparent container exposed to artificial light for six years prior to inert packaging and sterilization of the device.
Item Number 502001016 502004006 502004009 502004016 502010006
Worldwide Distribution-USA (nationwide) including DC and the states of AK, AL, AR, AZ, CA, CO, CT, FL, IL, IN, KS, LA, ME, MI, MO, MS, NC, ND, NE, NJ, NY, OH, OK, OR, PA, TN, TX, UT, VA, WA, and WI, and the countries of Canada, Australia, India, Austria, Japan, Czech Republic, Seoul Korea, France, Venezuela, Germany, Luxembourg, Spain, and Switzerland.
Voluntary: Firm initiated