Zimmer, Inc.: Medical Device Recall in 2014 - (Recall #: Z-1257-2014)
See the recall detail below. You can also see other recalls from the same firm in 2014.
Natural-Knee¿ II Posterior Stabilized (P.S.) Condylar Tibial Inserts: N-K II CONS PS TIB INS SZ N-K II CONS PS TIB INS SZ N-K II CONS PS TIB INS SZ N-K II CONS PS TIB INS SZ N-K II CONS PS TIB INS SZ N-K II CONS PS TIB INS SZ N-K II CONS PS TIB INS SZ N-K II CONS PS INS TIB SZ N-K II CONS PS TIB INS SZ N-K II CONS PS TIB INS SZ N-K II CONS PS TIB INS SZ N-K II CONS PS TIB INS SZ N-K II CONS PS TIB INS SZ N-K II CONS PS TIB INS SZ
Class II
A complaint was received for a discolored All Poly Tibia. Root cause analysis found that the discoloration was the result of free radical oxidation. The oxidation process was advanced prematurely as a result of the machined component being stored in a transparent container exposed to artificial light for six years prior to inert packaging and sterilization of the device.
Item Number 672509100 672509101 672509200 672509203 672511100 672511103 672513101 672513103 672513201 672516103 672519103 672522201 672525103 672525203
Worldwide Distribution-USA (nationwide) including DC and the states of AK, AL, AR, AZ, CA, CO, CT, FL, IL, IN, KS, LA, ME, MI, MO, MS, NC, ND, NE, NJ, NY, OH, OK, OR, PA, TN, TX, UT, VA, WA, and WI, and the countries of Canada, Australia, India, Austria, Japan, Czech Republic, Seoul Korea, France, Venezuela, Germany, Luxembourg, Spain, and Switzerland.
Voluntary: Firm initiated