Zimmer, Inc.: Medical Device Recall in 2014 - (Recall #: Z-1259-2014)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2014.

Data Source: FDA.
Product Description:

Apollo Knee Tibial Baseplate/Knee Tibial Insert: CONG TIB INS SZ0/19MM APO CONG TIB INS SZ0/22MM APO CONG TIB INS SZ1/9MM APOL CONG TIB INS SZ1/11MM APO CONG TIB INS SZ1/19MM APO CONG TIB INS SZ2/13MM APO CONG TIB INS SZ2/19MM APO CONG TIB INS SZ3/9MM APOL CONG TIB INS SZ3/13MM APO

Product Classification:

Class II

Date Initiated: February 25, 2014
Date Posted: March 26, 2014
Recall Number: Z-1259-2014
Event ID: 67503
Reason for Recall:

A complaint was received for a discolored All Poly Tibia. Root cause analysis found that the discoloration was the result of free radical oxidation. The oxidation process was advanced prematurely as a result of the machined component being stored in a transparent container exposed to artificial light for six years prior to inert packaging and sterilization of the device.

Status: Terminated
Product Quantity: 3638 units
Code Information:

Item Number 681100019 681100022 681101009 681101011 681101019 681102013 681102019 681103009 681103013

Distribution Pattern:

Worldwide Distribution-USA (nationwide) including DC and the states of AK, AL, AR, AZ, CA, CO, CT, FL, IL, IN, KS, LA, ME, MI, MO, MS, NC, ND, NE, NJ, NY, OH, OK, OR, PA, TN, TX, UT, VA, WA, and WI, and the countries of Canada, Australia, India, Austria, Japan, Czech Republic, Seoul Korea, France, Venezuela, Germany, Luxembourg, Spain, and Switzerland.

Voluntary or Mandated:

Voluntary: Firm initiated