Zimmer, Inc.: Medical Device Recall in 2014 - (Recall #: Z-1606-2014)
Updated on March 25, 2026.
See the recall detail below. You can also see other recalls from the same firm in 2014.
Data Source: FDA.
Product Description:
Tennis Elbow Support, Neoprene***LATEX FREE" Product Usage: Provides compression and warmth for mild to moderate elbow strains and sprains.
Product Classification:
Class II
Date Initiated: March 19, 2014
Date Posted: May 21, 2014
Recall Number: Z-1606-2014
Event ID: 68017
Reason for Recall:
During a transfer of products from a recently shutdown facility, the firm discovered raw material labeled as latex free actually contained latex.
Status: Terminated
Product Quantity: 649 units
Code Information:
Model #: 00-3448-003-00 through 00-3448-006-00
Distribution Pattern:
Worldwide Distribution - US Nationwide in the states of: AK, AZ, CA, FL, GA, IA, IL, IN, LA, MD, MI, MO NC, NE, NV, NY, OH, OK, OR, PA, RI, SD, TX, UT, VA, VT, WA, WI & WV and countries of: AUSTRALIA, CANADA, GERMANY, IRAQ, ITALY, SAUDI ARABIA, TUNISIA & UNITES ARAB EMIRATES.
Voluntary or Mandated:
Voluntary: Firm initiated