Zimmer, Inc.: Medical Device Recall in 2014 - (Recall #: Z-1842-2014)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2014.

Data Source: FDA.
Product Description:

VerSys¿ Hip System, Beaded Fullcoat Stems, Zimmer, Inc., Warsaw, IN 46581

Product Classification:

Class II

Date Initiated: May 19, 2014
Date Posted: July 2, 2014
Recall Number: Z-1842-2014
Event ID: 68339
Reason for Recall:

During routine inspection, beaded hip stem forgings received from supplier failed to meet ZES 2A-102. Processing of Beaded CoCrMo Alloy Hip Stem, material requirements per ATS #14-04818 and ATS #14-05549, respectively.

Status: Terminated
Product Quantity: 147 units
Code Information:

Item # 784301108, Lot #'s: 62478967, 62479742, 62503408, 62528282, 62530786, 62544915, 62554360; Item # 784301126, Lot #: 62544885, 62547211, Item # 784301156, Lot #, 62587079; Item # 784301208, Lot #'s: 62503411, 62509385, 62530790, 62547217, 62554362; Item # 784301226, Lot #'s: 62515502, 62547216; Item # 784301256, Lot #'s: 62489572, 62522386; Item # 784301326, Lot #, 62496464, 62594428, 62622893; Item # 784301356, Lot #: 62622894; Item # 784301382, Lot #: 62544907; Item # 784301511, Lot #: 62472423 & Item # 784301581, Lot #: 62496469.

Distribution Pattern:

Worldwide Distribution-USA (nationwide) including the states of AL, AZ, CA, CO, CT, FL, GA, IL, IN, MA, MI, MN, NC, ND, NJ, NY, OH, OK, PA, TN, TX, VA & WI, and the countries of Canada, Colombia, Dominican Republic, Japan, Korea (S. Korea), Peru, Taiwan & Thailand.

Voluntary or Mandated:

Voluntary: Firm initiated