Zimmer, Inc.: Medical Device Recall in 2014 - (Recall #: Z-2131-2014)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2014.

Data Source: FDA.
Product Description:

Zimmer Periarticular Locking Plate System Distal Radial Volar Plate Lateral Column (1st generation stainless steel only). Orthopedic fixation plate. For temporary internal fixation and stabilization of osteotomies and fractures.

Product Classification:

Class II

Date Initiated: June 10, 2014
Date Posted: August 6, 2014
Recall Number: Z-2131-2014
Event ID: 68657
Reason for Recall:

A complaint was received in which it was reported that an 8-hole plate from lot 62276042 had an enlarged hole that is unable to lock the screw, which resulted in the screw head pulling through plate hole.

Status: Terminated
Product Quantity: 28 units
Code Information:

Part Number 00-2358-014-04; lot 62276042

Distribution Pattern:

Worldwide Distribution - US including the states of CA and OH., and the countries of BRAZIL, CHINA, KOREA and TAIWAN.

Voluntary or Mandated:

Voluntary: Firm initiated