Zimmer, Inc.: Medical Device Recall in 2014 - (Recall #: Z-2548-2014)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2014.

Data Source: FDA.
Product Description:

Endo & Unipolar Femoral Heads: ENDO FEMORAL HEAD 43MM DIA, ENDO FEMORAL HEAD 53MM DIA, ENDO FEMORAL HEAD 47MM DIA, ENDO FEMORAL HEAD 50MM DIA, ENDO FEMORAL HEAD 42MM DIA, ENDO FEMORAL HEAD 54MM DIA, ENDO FEMORAL HEAD 48MM DIA, ENDO FEMORAL HEAD 41MM DIA, ENDO FEMORAL HEAD 42MM DIA, ENDO FEMORAL HEAD 43MM DIA, ENDO FEMORAL HEAD 44MM DIA, ENDO FEMORAL HEAD 45MM DIA, ENDO FEMORAL HEAD 46MM DIA, ENDO FEMORAL HEAD 47MM DIA, ENDO FEMORAL HEAD 48MM DIA, ENDO FEMORAL HEAD 49MM DIA, ENDO FEMORAL HEAD 50MM DIA, ENDO FEMORAL HEAD 51MM DIA, ENDO FEMORAL HEAD 52MM DIA, ENDO FEMORAL HEAD 53MM DIA, ENDO FEMORAL HEAD 54MM DIA, ENDO FEMORAL HEAD 55MM DIA

Product Classification:

Class II

Date Initiated: July 22, 2014
Date Posted: September 3, 2014
Recall Number: Z-2548-2014
Event ID: 68818
Reason for Recall:

Zimmer is recalling sterile-packaged implants and instruments packaged in six different package configurations due to packaging design verification test failures. Specifically, multiple test samples from each of the six configurations failed simulated distribution and shipping testing when the devices either breached inner or outer package trays or seal integrity was lost.

Status: Terminated
Product Quantity: 972,125 units
Code Information:

Item Numbers : 00781804300, 00781805300, 00781804700, 00781805000, 00781804200, 00781805400, 00781804800, 00781804100, 00781804200, 00781804300, 00781804400, 00781804500, 00781804600, 00781804700, 00781804800, 00781804900, 00781805000, 00781805100, 00781805200, 00781805300, 00781805400, 00781805500 Lot Numbers: 37210958, 37210967, 37211180, 37211183, 37211311, 37211316, 37211320, 37211379, 37211482, 62561484, 37211643, 37211645, 37211646, 37211647, 37211653, 37211655, 37211657, 37211658, 37211661, 37211801, 37211804, 37211805, 37211806, 37211811, 37211812, 37211815, 37211817, 37211820, 37212062, 37212063, 37212065, 37212074, 37212079, 62573574, 62573576, 62576002, 62576008, 62593298, 62593302, 37212608, 37211644

Distribution Pattern:

Worldwide Distribution.

Voluntary or Mandated:

Voluntary: Firm initiated