Zimmer, Inc.: Medical Device Recall in 2024 - (Recall #: Z-1720-2024)
See the recall detail below. You can also see other recalls from the same firm in 2024.
ZPLP Distal Lateral Fibular Plate, Left, 14 Holes, 210 MM, Sterile- Indicated for temporary internal fixation and stabilization of osteotomies and fractures Item Number: 47-2357-018-14
Class II
Thread form issue for the locking holes in that the locking screws may not properly mate with the plate. The improperly mated screw may not be readily recognizable by the user since the screw may not correctly lock, may result in delay in surgery or loss of fixation resulting further intervention
UDI: (01)00889024088610(17)290730(10)64381072 (01)00889024088610(17)300208(10)64661553 Lot Numbers: 64381072 64661553
Worldwide distribution - US Nationwide and the countries of Brazil, Canada, China, Hong Kong, India, Japan, New Zealand, Seoul Korea, Singapore, Switzerland, Taiwan.
Voluntary: Firm initiated