Zimmer Trabecular Metal Technology, Inc.: Medical Device Recall in 2019 - (Recall #: Z-1006-2019)
See the recall detail below. You can also see other recalls from the same firm in 2019.
NexGenTrabecular Metal Tibial Half Block Augment, Left Lateral/Right Medial, Size 4 5MM Thickness W/Screws Item Number: 00544800436 Product Usage: The Trabecular Metal Tibial Cone Augments are intended for use where severe degeneration, trauma, or other pathology of the knee joint indicates total knee arthroplasty. When used with the NexGen Complete Knee Solution Rotating Hinge Knee (RHK) System, the Trabecular Metal Tibial Cone Augments are for cemented use only. When used with the NexGen Complete Knee Solution Legacy Constrained Condylar Knee System, the Trabecular Metal Tibial Cone Augments are for cementless or cemented use.
Class II
Potentially comingled, resulting in the product in the box potentially not matching the product on the label
Lot Numbers: 63244259 63247617 63279968 61581585T (01) 00889024307902 (17) 210131 (10) 63244259 (01) 00889024307902 (17) 210228 (10) 63247617 (01) 00889024307902 (17) 210228 (10) 63279968 (01) 00889024307902 (17) 210228 (10) 61581585T
Worldwide Distribution - US Nationwide Distribution in the states of AL, CA, CO, ID, MS,NE, NJ, NY, OH, OR, PA, TX, VA, WI and Foreign CANADA, JAPAN, GERMANY , NETHERLANDS
Voluntary: Firm initiated