Nucletron BV: Medical Device Recalls in 2020

Updated on March 25, 2026.

According to to data from the FDA, there was 1 medical device recall made by Nucletron BV in 2020. See the detail of the recall below.

You can see similar recalls for other firms.

Data Source: FDA.
See this for other years:
2017
2020
  • There is a complaint regarding internal check cable guide mechanism failure for the Nucletron Flexitron brachytherapy afterloader system. Due to an obstruction during the check cable out drive, the internal guiding tube became displaced and the check cable became damaged. This issue could cause treatment interruption or in a very unlikely scenario, incorrect source positioning and radiation treatment errors.